Cardiac resynchronisation therapy as a cause of the electrical storm]

KARDIOLOGIA POLSKA(2007)

Cited 23|Views1
No score
Abstract
We report a case of young patient with dilated cardiomyopathy and implanted cardioverter-defibrillator in which resynchronisation therapy (CRT-D) induced an electrical storm. One month after implantation of a cardiac resynchronisation pacemaker (CRT-P) the patient suffered from ventricular tachycardia with poor haemodynamic status and was treated by implantation of a CRT-D with a Y adaptor. After replacement of the CRT-D due to Y adaptor damage (new device without a Y adaptor) we observed an electrical storm during ventricular pacing (biventricular, right and left ventricular pacing respectively). Changing pacing mode from DDDR to AAIR resolved ventricular tachycardias in that patient.
More
Translated text
Key words
cardiac resynchronisation therapy,electrical storm,implantable cardioverter-defibrillator