Automatic Implantable Cardioverter Defibrillator - Inadvertent Discharges During Permanent Pacemaker Magnet Tests

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(1987)

引用 33|浏览4
暂无评分
摘要
A patient with an automatic implantable cardioverter defibrillator (AICD) received two inadvertent shocks when a magnet was placed over the pacer during a routine permanent pacer check. Analysis of the rhythm strip suggested that both patients' QRS complexes (133 beats/minute) and asynchronous pacer artifacts (70 beats/minute) were counted by the AICD sensing system and exceeded the rate criteria of 153 beats/minute. This resulted in shocks from the AICD during sinus rhythm at 133 beats/minute. To avoid possible inadvertent shocks, an AICD should be deactivated while a magnet is placed over the pacemaker during a permanent pacer check.