Semiquantitative Measurement of Drugs in Urine Samples and Measurement of Pharmaceutical Active Substances in Serum Samples Using the Beckman Coulter Synchron LX20 PRO Clinical System.

Clinical Laboratory(2007)

Cited 23|Views2
No score
Abstract
Our manuscript describes the investigation of options for the detection and semiquantitative assay of drug (metabolites) and the assay of relevant pharmaceutical active substances using the LX20 PRO (Beckman Coulter) in comparison with the Dimension RxL (Dade Behring) and the TDx or AxSYM (Abbott) system. Drug tests carried out with the LX system satisfy sensitivity and selectivity requirements. The assay of the pharmaceutical active substances revealed a few noticeable clusters of values with valproate and, in particular, digitoxin, that can be attributable to the reagents used rather than to the system itself, which is highly stable and reproducible in terms of operation. These values are thus devoid of any considerable clinical significance.