Design of and data reduction from compact Thomson parabola spectrometers.

REVIEW OF SCIENTIFIC INSTRUMENTS(2011)

引用 25|浏览3
暂无评分
摘要
Thomson parabola spectrometers are used to characterize MeV ion beams produced in high intensity laser interactions. These spectrometers disperse multiple ion species according to their charge to mass ratio through the use of parallel electric and magnetic fields. Analytical solutions for ion deflection in electric and magnetic fields have been used to extract ion spectra with the assumption that fringing effects are negligible. Experimental space restrictions and dynamic range requirements necessitate designs that stress the analytical assumptions. Depending on design parameters, the error in the analytical assumption can be comparable to the energy resolution. Estimates are provided to approximate the error on the total ion deflection. A method for modeling ion trajectories including fringing effects is presented using software freely available or in common use. The magnetostatic fields are modeled in 3D, including material properties of nearby magnetic materials using RADIA. Electrostatic fields are modeled in 2D for a spectrometer implementing angled plates using the partial differential equation toolbox in MATLAB (R). Using these models to calculate the ion trajectory allows for analysis of a Thomson parabola spectrometer with an arbitrary field configuration. (C) 2011 American Institute of Physics. [doi: 10.1063/1.3556444]
更多
查看译文
关键词
magnetic materials,material properties,dynamic range,partial differential equation,analytic solution,data reduction,ion beam