Determination of roxatidine in human plasma by liquid chromatography/electrospray mass spectrometry and application to a clinical pharmacokinetic study.

RAPID COMMUNICATIONS IN MASS SPECTROMETRY(2007)

Cited 9|Views9
No score
Abstract
A rapid and sensitive liquid chromatography/mass spectrometry (LC/MS) method was developed and validated for the determination of roxatidine in human plasma. Roxatidine was extracted by single liquid-liquid extraction with tert-butyl methyl ether, and the chromatographic separation was performed on a C-8 column. The total analytical run time was relatively short (5 min), and the limit of assay quantification was 2 ng/mL using 0.1 mL of human plasma. Roxatidine and the internal standard, propranolol, were monitored in selected ion monitoring (SIM) mode at m/z 307.3 and 260.3, respectively. The standard curve was linear over a concentration range from 2-500 ng/mL, and the correlation coefficients were > 0.999. The mean intra- and inter-day assay accuracy ranged from 103.4-108.8% and 102.3-110.0%, respectively, and the mean intra- and inter-day precision was between 3.3-8.8% and 5.3-6.2%, respectively. The developed assay method was successfully applied to a pharmacokinetic study in human volunteers after oral administration of roxatidine acetate hydrochloride at a dose of 75 mg. Copyright (c) 2007 John Wiley & Sons, Ltd.