Single-chamber, rate-responsive pacemaker-mediated tachycardia.

Canadian Journal of Cardiology(2010)

Cited 2|Views1
No score
Abstract
The case of a patient with a unipolar VVIR, who developed a sensor-driven increased pacing rate at rest following pulse generator replacement, is presented. The cause was pectoral muscle stimulation, which triggered a sensor-driven rate response in the supine position. The possible causes and management are discussed.
More
Translated text
Key words
Pectoral muscle stimulation,Rate responsive pacing
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined