Single-chamber, rate-responsive pacemaker-mediated tachycardia.
Canadian Journal of Cardiology(2010)
Abstract
The case of a patient with a unipolar VVIR, who developed a sensor-driven increased pacing rate at rest following pulse generator replacement, is presented. The cause was pectoral muscle stimulation, which triggered a sensor-driven rate response in the supine position. The possible causes and management are discussed.
MoreTranslated text
Key words
Pectoral muscle stimulation,Rate responsive pacing