[Ventilation of an unprotected airway: evaluation of a new peak-inspiratory-flow and airway-pressure-limiting bag-valve-mask].
Der Anaesthesist(2006)
摘要
In this model of a simulated unprotected airway, a reduction of inspiratory time from 2 s to 1 s using the Smart-Bag resulted in comparable inspiratory peak airway pressure and lower, but clinically comparable, lung tidal volume. Stomach inflation occurred only at a LOSP of 0.49 kPa (5 cm H2O), and was higher with an inspiratory time of 2 s vs 1 s.