Chrome Extension
WeChat Mini Program
Use on ChatGLM

Rarebit Perimetry For Bedside Testing: Comparison With Standard Automated Perimetry

JOURNAL OF NEURO-OPHTHALMOLOGY(2010)

Cited 4|Views5
No score
Abstract
Background: Rarebit perimetry (RBP), a technique developed for the detection of early damage to the afferent visual system, has not been extensively tested at the bedside. This study was designed to test the feasibility of bedside testing with RBP in comparison with standard automated perimetry (SAP) performed in the clinic.Methods: We tested 29 eyes of 15 subjects admitted with neurologic or neurosurgical diseases affecting the afferent visual system. RBP was performed on a laptop computer at the bedside. SAP (Humphrey field analyzer) testing was performed later in the clinic.Results were evaluated by a masked neuro-ophthalmologist. Results: Visual fields corresponded between RBP and SAP in 21(72%) of the 29 tested eyes. RBP detected defects in 5 subjects who had normal visual field results on SAP. All subjects preferred RBP for convenience.Conclusion: RBP is a convenient method of bedside visual field testing and is no less sensitive to visual field defects in this role than SAP. Journal of Neuro-Ophthalmology 2010;30:243-247 doi: 10.1097/WNO.0b013e3181dee88a (C) 2010 by North American Neuro-Ophthalmology Society
More
Translated text
Key words
rarebit perimetry,bedside testing,automated perimetry