Error correction algorithm for high accuracy bio-impedance measurement in wearable healthcare applications

Circuits and Systems(2013)

引用 4|浏览0
暂无评分
摘要
High accuracy in measurement of physiological signals like skin impedance is critical and is more difficult to achieve in biomedical systems with very limited power budget. A new method for accurate bio-impedance measurement using square wave demodulation is proposed in this paper. An error correction algorithm has been derived from the analysis of a generic impedance model with multiple poles, which reduces the estimation error in square wave demodulation from 9.3% to 1.3% for a simple tissue model. Simulation results in Matlab using ideal RC values show an accuracy of <;0.14% for single pole and <;0.67% for two pole RC networks. Measurements from saline phantom solution gives an accuracy <;0.72%. The results indicate that the algorithm can be used as a postprocessing technique for error correction or even incorporated into wearable signal monitoring SoCs since modulation and demodulation using square waves consumes less power compared to sine waves.
更多
查看译文
关键词
biomedical electronics,biomedical measurement,error correction,health care,medical signal processing,patient care,system-on-chip,Matlab,biomedical systems,error correction algorithm,estimation error reduction,generic impedance model-multiple-poles,high-accuracy bioimpedance measurement,ideal RC values,physiological signal measurement,saline phantom solution,skin impedance,square wave demodulation,tissue model,two-pole RC networks,wearable healthcare applications,wearable signal monitoring SoCs,Bio-impedance,Physiological signal monitoring,