A New Transducer For 3d Harmonic Imaging

Ultrasonics Symposium, 2002. Proceedings. 2002 IEEE(2002)

引用 20|浏览1
暂无评分
摘要
We developed a fast rotating ultrasound transducer for 3D-echocardiography containing a broadband linear array. In the harmonic mode the tissue to clutter ratio was measured as function of frequency. This resulted in an optimal harmonic transmit frequency of 2.0 MHz. In-vitro evaluation, using the optimal frequency, showed an average volume error, for 5 reconstructed agar phantoms, of approximately 1%. Further, a clinical case study proved the feasibility of dynamic left ventricular volume measurements for the transducer.
更多
查看译文
关键词
biomedical transducers,echocardiography,ultrasonic transducer arrays,2.0 MHz,3D harmonic imaging,3D-echocardiography,average volume error,broadband linear array,dynamic left ventricular volume measurements,fast rotating ultrasound transducer,harmonic mode,optimal harmonic transmit frequency,reconstructed agar phantoms,tissue to clutter ratio,transducer,