A method to determine diastolic blood pressure based on pressure pulse propagation in the electronic palpation method

Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE  (2001)

引用 8|浏览3
暂无评分
摘要
Evaluates the accuracy of the pulse transit time method for determining diastolic blood pressure using intra-arterial blood pressure as a reference. First, the paper describes the method. Then it discusses two sets of measurements which were carried out to determine the accuracy of the method with cardiac operated patients and healthy young volunteers. The thus obtained accuracies were +0.7 mmHg ± 10.7 mmHg for the cardiac patients and -6.6 mmHg ± 10.5 mmHg for the young volunteers, respectively.
更多
查看译文
关键词
biomedical transducers,blood pressure measurement,patient monitoring,pressure transducers,accuracy,cardiac operated patients,diastolic blood pressure,electronic palpation method,healthy young volunteers,intra-arterial blood pressure,multi-element transducer array,pressure pulse propagation,pulse transit time method,blood pressure,testing,pressure measurement,transducers