Implanted electrodes for multi-month EEG.

EMBC(2014)

引用 3|浏览15
暂无评分
摘要
An implanted electroencephalogram (EEG) recorder would help diagnose infrequent seizure-like events. A proof-of-concept study quantified the electrical characteristics of the electrodes planned for the proposed recorder. The electrodes were implanted in an ovine model for eight weeks. Electrode impedance was less than 800 Ohms throughout the study. A frequency-domain determination of sedation performed similarly for surface versus implanted electrodes throughout the study. The time-domain correlation between an implanted electrode and a surface electrode was almost as high as between two surface electrodes (0.86 versus 0.92). EEG-certified clinicians judged that the implanted electrode quality was adequate to excellent and that the implanted electrodes provided the same clinical information as surface electrodes except for a noticeable amplitude difference. No significant issues were found that would stop development of the EEG recorder.
更多
查看译文
关键词
resistance 800 ohm,medical disorders,neurophysiology,biomedical electrodes,eeg-certified clinicians,electroencephalography,sedation,amplitude difference,frequency-domain analysis,clinical information,prosthetics,proof-of-concept study,electrode impedance,infrequent seizure-like events,implanted electrode quality,time-domain analysis,implanted electroencephalogram recorder,frequency-domain determination,ovine model,surface electrode,electrode electrical characteristics,multimonth eeg,time-domain correlation,eeg recorder