Noncontact ECG recording system with real time capacitance measurement for motion artifact reduction.

IEEE Trans. Biomed. Circuits and Systems(2014)

引用 42|浏览29
暂无评分
摘要
A system for noncontact ECG recording is proposed that measures the real time electrode-body capacitance concurrently with the ECG as a reference signal for motion artifact reduction. Simultaneous recordings of these two signals from the human body in the presence of electrode motion artifacts are shown and an adaptive least-mean-squares (LMS) filtering algorithm run on these signals is demonstrated to be able to reduce the severity of certain types of electrode motion artifacts.
更多
查看译文
关键词
electrocardiography,bioelectric potentials,motion artifact reduction,capacitance measurement,real time electrode-body capacitance measurement,biomedical electrodes,capacitive sensors,capacitance,medical signal processing,noncontact ecg recording system,least mean squares methods,adaptive filters,wearable sensors,adaptive least-mean-squares filtering algorithm