Coplanar Electrode Layout Optimized For Increased Sensitivity For Electrical Impedance Spectroscopy

MICROMACHINES(2015)

Cited 38|Views49
No score
Abstract
This work describes an improvement in the layout of coplanar electrodes for electrical impedance spectroscopy. We have developed, fabricated, and tested an improved electrode layout, which improves the sensitivity of an impedance flow cytometry chip. The improved chip was experimentally tested and compared to a chip with a conventional electrode layout. The improved chip was able to discriminate 0.5 mu m beads from 1 mu m as opposed to the conventional chip. Furthermore, finite element modeling was used to simulate the improvements in electrical field density and uniformity between the electrodes of the new electrode layout. Good agreement was observed between the model and the obtained experimental results.
More
Translated text
Key words
electrical impedance spectroscopy,sensitivity,improved design,impedance flow cytometry