[Measurement of central corneal thickness in myopic eyes with ultrasound and Pentacam scheimpflug system].

[Zhonghua yan ke za zhi] Chinese journal of ophthalmology(2006)

引用 15|浏览8
暂无评分
摘要
OBJECTIVE:To assess the accuracy and variability of central corneal thickness measurements in myopic eyes obtained by ultrasound pachymetry and the Pentacam scheimpflug imaging system. METHODS:First, corneal thickness in 50 healthy eyes was measured three times with the Pentacam system to evaluate the repeatability. Then, Scheimpflug imaging system (Pentacam) and A-scan ultrasonic pachymeter (SP-3000) were used to measure the corneal thickness on 217 myopic cases (433 eyes) undergoing laser in situ keratomileusis (LASIK) and 63 cases (126 eyes) 6 months after LASIK. The results were compared. RESULTS:Repeated-measures ANOVA showed the repeatability of the Pentacam measurements was high (P > 0.05). Pre-operatively, the value of corneal central thickness measured by ultrasonic pachymetry and the Pentacam system was (537.26 +/- 31.61) microm and (538.08 +/- 32.72) microm, respectively. There was no significant difference between the results of these two methods (t = 1.867, P > 0.05). The difference between these two methods ranged -35 - 34 microm, and was within -20 - 20 microm in 95.38% cases. Post-operatively, the value of corneal central thickness measured by ultrasonic pachymetry and the Pentacam system was (464.21 +/- 41.04) microm and (465.17 +/- 42.65) microm, respectively. There was no significant difference between these two methods, too (t = 0.913, P > 0.05). There was a significant linear correlation between the two methods (P < 0.001). The thinnest point of cornea was mostly located surrounding the apex and inclined towards inferotemporal. CONCLUSION:The Pentacam scheimpflug system offers an acceptable method for evaluating corneal pachymetry, which is useful in corneal refractive surgery.
更多
查看译文
关键词
Cornea,Tomography, optical coherence,Ultrasonography
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要