We propose a general global optimal algorithm to optimize the miniature permanent magnet structure of a micro magnetic resonance chip (mu NMR-chip). For this purpose, we analyze the sensitivity of the permanent magnet structure to the design variables and determine the optimization variables. After this, radial basis function neural networks (RBFNNs) are constructed to model the objective functions, and the nondominated sorting genetic algorithm II (NSGA II) is improved by introducing a different weighting factor for each objective function in calculating the crowding distance. Combining the RBFNN with the improved NSGA II optimizes the miniature permanent magnet structure. Through comparison, the optimization solutions are proven effective. Finally, the optimized permanent magnet structure is manufactured and tested experimentally. After optimization, the volume of the permanent magnet block is reduced by 39%, and the permanent magnet becomes easier to manufacture.
More
Translated text
Key words
Microfluidic chip, nuclear magnetic resonance, permanent magnet, multiobjective optimization