Determination Of Dioxane Residue In Cosmetics By Isotope Dillution-Headspace Gas Chromatography-Mass Spectrometry

CHINESE JOURNAL OF CHROMATOGRAPHY(2013)

Cited 7|Views12
No score
Abstract
A comprehensive analytical method based on isotope dilution-headspace gas chromatography-mass spectrometry was developed for the determination of dioxane residue in the cosmetics. Cosmetic sample (2 g) was placed in a headspace vial. Then, 1 g sodium chloride, 1 mL d(8)-dioxane solution (20 mg/L) and 9 mL water were added. The vial was sealed and heated at 70 degrees C for 40 min. The qualitative and quantitative analyses were carried out for the analyte in selected ion monitoring (SIM) mode after the chromatographic separation on an HP-5MS column (30 m x 0.25 mm x 0.25 mu m). The limit of quantification (LOQ) for dioxane was 2.5 mg/kg. The mean recoveries were 90.5% - 104.0% at the spiked levels of 2.5 - 50 mg/kg with the relative standard deviations in the range of 1.26% - 3.98%. The method is accurate, simple, rapid, and applicable for the inspection of dioxane residue in cosmetics.
More
Translated text
Key words
isotope dilution, headspace gas chromatography-mass spectrometry (HS-GC-MS), dioxane, cosmetics