A novel fixation system for sacroiliac dislocation fracture: internal fixation system design and biomechanics analysis.

Clinical Biomechanics(2013)

引用 9|浏览2
暂无评分
摘要
Although there were many different types of fixation techniques for sacroiliac dislocation fracture, the treat remained challenging in posterior pelvic ring injury. The purpose of this study was to evaluate the biomechanical effects of a novel fixation system we designed.12 human cadavers (L3-pelvic-femora) were used to compare biomechanical stability after reconstruction on the same specimens in four conditions: (1) intact, (2) cable system, (3) plate-pedicle screw system, and (4) cable system and plate-pedicle screw combination system (combination system). Biomechanical testing was performed on a material testing machine for evaluating the stiffness of the pelvic fixation construct in compression and torsion.The cable system and plate-pedicle screw system alone may be insufficient to resist vertical shearing and rotational loads; however the combination system for unstable sacroiliac dislocation fractures provided significantly greater stability than single plate-pedicle or cable fixation system.The novel fixation system for unstable sacroiliac dislocation fractures produced sufficient stability in axial compression and axial rotation test in type C pelvic ring injuries. It may also offer a better solution for sacroiliac dislocation fractures.