Negative Interference In Serum Hbsag Elisa From Rheumatoid Factors

PLOS ONE(2013)

引用 10|浏览3
暂无评分
摘要
Background: RF(Rheumatoid factor) is usually thought to cause positive interference in immunoassay. Recently, our study showed that high-concentration RFs caused negative interference as well as positive interference in serum HBsAg(Hepatitis B surface antigen) ELISA(Enzyme-linked immunosorbent assay), but it is unclear that RF causing negative interference is an anomaly produced by a certain ELISA kit or a common property of most of HBsAg ELISA kits.Methods: Serum models were made by blending HBsAg-positive sera and high-or moderate-concentration RFs sera at the ratio of 1:9, then one-step and two-step ELISA were adopted to determine HBsAg in serum models.Results: No matter what kind of kit used, one-step ELISA showed that HBsAg S/CO(sample/cut off) values in serum models were significantly lower than original values. Bivariate correlations tests showed decline rates of HBsAg S/CO Values were not associated to serum RF concentrations ranging from 288 to 3560 IU/mL. HBsAg converted to be negative in 69.80% serum models with original-value ranging from 1.00 to 10.00, and in 2.68% serum models with higher original-value. RF causing decline of HBsAg S/CO value provided by one-step ELISA was more obvious than that provided by two-step ELISA.Conclusions: It is concluded that susceptibility of all HBsAg ELISA assays to interference from RF, leading to predominantly lower and in some cases "false-negative" results, and moreover, the lower the original HBsAg S/CO Value, the higher the false-negative rate.