Differential pulse voltammetry method for the determination of glucose on multi-nanostructured active electrode modified by platinum nanoparticles

ADVANCES IN CHEMISTRY RESEARCH II, PTS 1-3(2012)

Cited 0|Views7
No score
Abstract
Electrocatalytic oxidation of multi-nanostructured active electrode modified by platinum nanoparticles on glucose was examined. Based on 0.5mol/L KOH solution, we see a sensitive and good-shaped oxidation peak current near -0.77V (VS, SCE) by using differential pulse voltammetry method to scan in the range of -0.9 similar to 0.4V.The peak is regarded as the quantitative peak. There is a good linear relationship between glucose concentration and the peak current in the range of 1.0x10(-2)similar to 1.0x10(-5) mol/L. The linear correlation coefficient is 0.99864.The detection limit is 1.0x10(-6)mol/L. If added 0.06 m mol/L ascorbic acid or 0.3m mol/L uric acid (simulating human blood components), the determination of glucose is not interfered with. Results of the determination of glucose concentration in the blood are satisfactory.
More
Translated text
Key words
Multi-nanostructured active electrode,Glucose,Electrocatalytic oxidation,Differential pulse voltammetry