Determination of trace impurities in high-purity zirconium oxide by high-resolution inductively coupled plasma mass spectrometry

BUNSEKI KAGAKU(2004)

引用 5|浏览1
暂无评分
摘要
Using high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS), trace impurities in high-purity zirconium oxide powder samples were determined. A 0.5 g zirconium oxide sample was dissolved with 5 in] of sulfuric acid (1 + 1) in a PTFE pressure vessel at 230 C for 40 h. Most of the spectral interferences were avoided by using a high-resolution spectrometer. The matrix effects of Zr were investigated. Since the matrix effects of a high Zr concentration on the peaks of the internal standard were similar to those on almost all of the analyte elements, except for Hf, the internal-standard method was employed for quantitative analysis. Indium was used for the internal standard element. The analytical values, except for Hf obtained by the internal-standard method, approximately agreed with those obtained by the standard addition method. The analytical values of the Hf obtained by ICP-AES approximately agreed with those obtained by the standard addition method. The detection limits in the solid samples we're in the range of 0.01similar to9 mug g(-1). The determinations for Na, Mg, Al, Ca; Ti, V, Cr, Mn, Fe, Ni, Sr, Cs, La, Ce, Hf, Pb and Bi in three kinds of commercially available high-purity zirconium oxide powders are presented.
更多
查看译文
关键词
high-resolution inductively coupled plasma mass spectrometry,high-purity zirconium oxide,matrix effects,internal standard method