Position resolution and efficiency measurements with large scale Thin Gap Chambers for the super LHC

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2011)

Cited 12|Views9
No score
Abstract
New developments in Thin Gap Chamber (TGC) detectors to provide fast trigger information and high precision muon tracking under sLHC conditions are presented. The modified detectors are shown to stand a high total irradiation dose equivalent to 6C/cm of wire, without showing any deterioration in their performance. Two large (1.2×0.8m2) prototypes containing four gaps, each gap providing pad, strip and wire readout, with a total thickness of 50mm, have been constructed. Their local spatial resolution has been measured in a 100GeV/c muon test beam at CERN. At perpendicular incidence angle, a single gap position resolution of better than 60μm has been obtained. For an incidence angle of 20° a resolution of less than 100μm was achieved. TGC prototypes were also tested under a flux of 105Hz/cm2 of 5.5–6.5MeV neutrons, showing a high efficiency for the detection of cosmic muons.
More
Translated text
Key words
Super LHC,Thin Gap Chamber,Aging effect,Position resolution