High-performance liquid chromatography analysis of anti-inflammatory pharmaceuticals with ultraviolet and electrospray-mass spectrometry detection in suspected counterfeit homeopathic medicinal products.

Journal of Pharmaceutical and Biomedical Analysis(2007)

引用 79|浏览26
暂无评分
摘要
A simple high-performance liquid chromatography (HPLC) method with both ultraviolet (UV) and electrospray ionisation mass spectrometry (ESI-MS) detection has been developed for the determination of seven pharmaceuticals in counterfeit homeopathic preparations. Naproxen, Ketoprofen, Ibuprofen, Diclofenac, Piroxicam, Nimesulide and Paracetamol were separated by reversed phase chromatography with acetonitrile–water (0.1% acetic acid) mobile phase, and detected by UV at 245nm and by ESI-MS in negative ionisation mode with the exception of Paracetamol which was detected in positive ionisation mode. Benzoic acid was used as internal standard (IS). This method was successfully applied to the analysis of homeopathic preparations like mother tinctures, solutions, tablets, granules, creams, and suppositories. Linearity was studied with UV detection in the 50–400μgmL−1 range and with ESI-MS in the 0.1–50μgmL−1 range. Good correlation coefficients were found in both UV and ESI-MS. Detection limits ranged from 0.18 to 41.5ng in UV and from 0.035 to 1.00ng in ESI-MS.
更多
查看译文
关键词
Homeopathy,Homeopathic preparations,NSAIDs,HPLC–UV,HPLC–ESI-MS