A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 2).

Biosensors and Bioelectronics(2003)

Cited 119|Views17
No score
Abstract
The aim of this study was to evaluate the reproducibility, the accuracy and the reliability of a continuous subcutaneous glucose measuring system. The GlucoDay® system (A. Menarini I.F.R. S.r.l.—Florence, Italy) is a portable instrument provided with a micro-pump and a biosensor, coupled to a microdialysis system (see part 1). This instrument has demonstrated high reliability coupled with a low degree of invasivity. The profiles of glucose monitoring allow to achieve an excellent knowledge of the real variation of glucose in diabetic patients. The reproducibility study showed a bias lower than 10% between instruments. The accuracy study showed a difference from the reference method lower than 15%.
More
Translated text
Key words
Biosensors,Continuous monitoring,Glucose,Microdialysis