A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 2).
Biosensors and Bioelectronics(2003)
Abstract
The aim of this study was to evaluate the reproducibility, the accuracy and the reliability of a continuous subcutaneous glucose measuring system. The GlucoDay® system (A. Menarini I.F.R. S.r.l.—Florence, Italy) is a portable instrument provided with a micro-pump and a biosensor, coupled to a microdialysis system (see part 1). This instrument has demonstrated high reliability coupled with a low degree of invasivity. The profiles of glucose monitoring allow to achieve an excellent knowledge of the real variation of glucose in diabetic patients. The reproducibility study showed a bias lower than 10% between instruments. The accuracy study showed a difference from the reference method lower than 15%.
MoreTranslated text
Key words
Biosensors,Continuous monitoring,Glucose,Microdialysis