Determination of etoposide in human plasma and leukemic cells by high-performance liquid chromatography with electrochemical detection.

Journal of Chromatography B: Biomedical Sciences and Applications(2001)

引用 9|浏览5
暂无评分
摘要
This paper describes a high-performance liquid chromatographic method with electrochemical detection for the determination of etoposide levels in plasma, total and non-protein bound concentration, and in leukemic cells. The precision for between-runs (n=6) was 7.0, 4.9, and 9.5%, the accuracy was 3.7, 7.1 and 6.3%, and within-runs precision (n=6) was 3.9, 2.9 and 5.1% for total plasma, non-protein bound plasma fraction and leukemic cells, respectively. The correlation coefficients (R2) were 1.00 for all calibration curves. These assays have been applied to analyze samples from one patient with acute myelogenous leukemia during 24 h after i.v. infusion of etoposide (100 mg/m2).