Chrome Extension
WeChat Mini Program
Use on ChatGLM

Evaluation of uncertainty of measurement from method validation data: An application to the simultaneous determination of retinol and α-tocopherol in human serum by HPLC

Journal of Chromatography B(2009)

Cited 29|Views7
No score
Abstract
An isocratic RP-HPLC method for the determination of retinol and α-tocopherol in serum, with fluorescence and UV/VIS detection, respectively, was developed and validated according to international guidelines. Detection (retinol 0.015mg/L, α-tocopherol 0.29mg/L) and quantification (retinol 0.05mg/L, α-tocopherol 0.95mg/L) limits were determined. Repeatability was <3.3% and <2.9% and intermediate precision was <4.6% and <3.2%, for retinol and α-tocopherol, respectively. Certified reference materials were utilised to assess bias and guarantee traceability to SI units. Expanded uncertainties (retinol 8.9%; α-tocopherol 7.9%), estimated according to the EURACHEM/CITAC guide from method validation data, satisfied fit-for-purpose requirements based on biological variability.
More
Translated text
Key words
Serum,Retinol,α-Tocopherol,HPLC,Validation,Uncertainty of measurement,Biological variability,Quality specifications
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined