Expanding of FOV of NaI(Tl) gamma camera detectors—Is it possible?

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2006)

引用 5|浏览8
暂无评分
摘要
Every gamma camera detector used for medical imaging of conventional design faces the problem of distorted or no information readout at the areas closer to the edge of detectors. Obtaining position and energy information becomes almost impossible at distance 0–12 of PMT size from the edge. Therefore, in some designs were proposed losing of edge energy resolution while improving in spatial uniformity when it comes to imaging at the edges. This work is dedicated to understanding of the problem, mathematical simulations, practical tests and recommendations to build detectors with larger usable FOV without increasing in dimensions. To study the problem we built the test jig with linear motion source and readout electronics to simulate gamma cameras of PMTs. Based on simulation results the idea of modifying of crystal shape combined with specific light redirection system of baffles was tested and allowed to expand usable FOV. The results are presented and showed that for traditional NaI(Tl) scintillators using 2″ PMT may be possible to obtain relatively good spatial resolution starting from 4–5mm from the edge of a detector. The question of economical efficiency of proposed method is being investigated and a special detector manufacturing technology must be developed to accommodate this. While we believe that achieved results are very important for small size detectors (<20cm) they could be beneficial even for larger detectors used in whole body imaging systems.
更多
查看译文
关键词
Gamma camera,Expanding,FOV
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要