Intercomparison of characterization techniques of filter radiometers in the ultraviolet region

METROLOGIA(2003)

引用 7|浏览11
暂无评分
摘要
Narrow-band filter radiometers at 248 nm, 313 nm, 330 nm and 368 nm wavelengths were used to compare calibration facilities of spectral (irradiance) responsivity at HUT, NPL and BNM-INM. The results are partly in agreement within the stated uncertainties. Use of demanding artefacts in the intercomparison revealed that the wavelength scales of the participating institutes deviate more than expected. Such effects cannot be seen in typical intercomparisons of spectral responsivity or spectral transmittance, where spectrally neutral samples are used.