Absorption Characteristics At 1.9 Mu-M - Effect On Vascular Welding

LASERS IN SURGERY AND MEDICINE(1993)

引用 48|浏览13
暂无评分
摘要
A 1.9 mum laser was used to investigate the acute weld strengths for anastomoses of rat and rabbit aortas and femoral arteries. The wall thicknesses for these vessels approximately matched the optical absorption depth of 125 mum for 1.9 mum radiation in vascular tissues. A low power (150 mW) 1.9 mum laser was used. Laser power was delivered through silica fiber optics for manual control. The fiber tip was held approximately 1 mm from the, target resulting in a laser spot size of 0.7 mm at the tissue. The linear delivery rate was approximately 0.3 mm/sec. Acute burst pressures of the welds showed a linear correlation with the reciprocal of the vessel radius. These results suggest that the product of the weld strength times the optical absorption depth is constant over the range of vessel sizes studied. A weld strength for a weld thickness equal to the optical absorption depth was determined to be 4 x 10(6) dynes/cm2, which is comparable to the strength of sutured anastomoses. These acute studies suggest that a laser wavelength with absorption depth in tissue matched to the vessel wall thickness should yield optimum welds. Therefore, a laser operating near 1.9 mum is suitable for small vessel welding.
更多
查看译文
关键词
BURST STRENGTH, OPTICAL PENETRATION DEPTH, VESSEL WALL THICKNESS