Optimizing the geometry of implantable leads for recording the monophasic action potential with fractally coated electrodes.

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(1998)

引用 12|浏览4
暂无评分
摘要
This study investigates the influence of various lead geometry on intracardial signals like the monophasic action potential (MAP) to optimize the geometry of implantable MAP leads The experimental results were compared with afield theoretical approach to the origin of MAP from the transmembrane potential (TAP). During the experiments several lead geometries (tip surface. 1.3 to 12 mm(2); tip-ring distance: 0.8 mm to 25 cm; ring surface: 1.8mm(2) to 40 mm(2)) were investigated in endo- and epicardial positions in 12 dogs (17+/-9 kg). The electrodes were fixed passively (tines) or actively (screws). MAP was recorded during several interventions and correlated with MAP measured using an Ag-AgCl MAP catheter. The experimental results showed that small tips provided high MAP amplitudes with less pressure. No difference was observed using active and passive fixations. A tip-ring distance smaller than 5 mm with a ring surface smaller than the tip (<5 mm(2)) avoided artifacts in the repolarization course. For the theoretical approach the quasistatic, anisotropic bidomain model was calculated in small unity volumes V-i,where the TAP phi(m)(i), was constant and represented by the current density J(i). Two solutions for electrode positions at anal outside the heart were achieved. By superposition of each solution phi(ei) the summed potential at the electrode position was calculated The theoretical findings show in good correlation with the experimental results that a larger distance than 10 mm leads to distortions in repolarization course by signals proportional to phi(out).
更多
查看译文
关键词
monophasic action potential,leads,action potential,field theory
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要