Time-resolved immunofluorometric assay for thyrotropin improved for enhanced functional sensitivity and clinical discrimination.

SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION(2009)

引用 5|浏览5
暂无评分
摘要
Modification of a new commercial time-resolved immunofluorometric assay for thyrotropin (Delfia(R) hTSH Ultra) by prediluting samples, reducing tracer concentration, and increasing total reaction volume offered improved functional sensitivity of 0.005 mIUl(-1) (20% between-assay CV) over the functional sensitivity limit of 0.014 mIUl(-1) of the unmodified TSH assay. Serum TSH concentrations of 50 hyperthyroid patients and 24 clinically euthyroid patients with serum TSH < 1.1 mIUl(-1) by a second generation immunoradiometric assay were analysed by both the original and modified TSH-IFMA method. The discrimination was better with the modified method, according to ROC-analysis (p = 0.027). In conclusion, these modifications enhanced functional sensitivity and improved the discriminatory power of the assay in separating hyperthyroid patients from euthyroid subjects with low TSH concentrations.