A-150 plastic radiometric calorimeter for charged particles and other radiations

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(1986)

引用 11|浏览14
暂无评分
摘要
A local absorbed dose calorimeter with certain novel features has been designed and constructed of A-150 tissue-equivalent (TE) plastic. The radiation absorption properties of this material and the relative absence of core impurities make the calorimeter suitable for use in a wide variety of radiation beams. The low thermal diffusivity of A-150 plastic led to the development of a unique spiral electrical calibration heater that has a mass of only 0.1% that of the core. The calorimeter can be calibrated in the quasiadiabatic or the heat-loss-compensated mode to test for possible effects caused by temperature gradients. The details of construction and operation are described.
更多
查看译文
关键词
heat loss,thermal diffusivity,charged particles,temperature gradient