Simultaneous reconstruction of scintillation light and ionization charge produced by 511keV photons in liquid xenon: Potential application to PET

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2009)

Cited 18|Views13
No score
Abstract
In order to assess the performance of liquid xenon detectors for use in positron emission tomography we studied the scintillation light and ionization charge produced by 511keV photons in a small prototype detector. Scintillation light was detected with large area avalanche photodiodes while ionization electrons were collected on an anode instrumented with low noise electronics after drifting up to 3cm. Operational conditions were studied as a function of the electric field. Energy resolutions of <10% (FWHM) were achieved by combining the scintillation light and ionization charge signals. The relationship between scintillation light and ionization signals was investigated. An analysis of the sources of fluctuations was performed in order to optimize future detector designs.
More
Translated text
Key words
Liquid xenon,PET,Medical imaging,TPC