Improvement of an enzyme electrode by poly(vinyl alcohol) coating for amperometric measurement of phenol.

Talanta(2007)

引用 14|浏览6
暂无评分
摘要
A poly(vinyl alcohol) film cross-linked with glutaraldehyde (PVA–GA) was introduced to the surface of a tyrosinase-based carbon paste electrode. The coated PVA–GA film was beneficial in terms of increasing the stability and reproducibility of the enzyme electrode. The electrode showed a sensitive current response to the reduction of the o-quinone, which was the oxidation product of phenol, by the tyrosinase, in the presence of oxygen. The effects of the PVA and PVA–GA coating, the pH, and the GA:PVA ratio on the current response were investigated. The sensitivity of the PVA–GA–Tyr electrode was 130.56μA/mM (1.8μA/μMcm2) and the linear range of phenol was 0.5–100μM. At a higher concentration of phenol (>100μM), the current response showed the Michaelis–Menten behavior. Using the PVA–GA–Tyr electrode, a two-electrode system was tested as a prototype sensor for portable applications.
更多
查看译文
关键词
Tyrosinase,Poly(vinyl alcohol),Glutaraldehyde,Carbon paste electrode,Phenol
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要