Robot-Based Nde System Using 3d-Mobile Hts-Squid

IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY(2009)

Cited 8|Views26
No score
Abstract
A novel HTS-SQUID gradiometer with ramp-edge Josephson junctions (JJs) composed of SmBa2Cu3Oy base-electrode and La01Er0.95Ba1.95Cu3Oy counter-electrode was developed. The robustness of the gradiometer in AC field perpendicular to the device plane was investigated. In an AC field of less than 2.8 mT(pp) at 100 Hz, flux-trapping, jumping and increase in noise did not happen in the gradiometer, while a traditional HTS-SQUID gradiometer with YBa2Cu3O7-x thin film and bicrystal JJs with the same dimension as the novel gradiometer could not be operated properly in an AC field of about several tens nT(pp). A robot-based NDE system utilizing the novel HTS-SQUID gradiometer was developed. The system could move the gradiometer three-dimensionally (3D) in an unshielded environment without magnetic shielding or compensation technique. Eddy-current-based detection of a through hole in magnetized stainless steel board was demonstrated by the robot-based HTS-SQUID NDE system.
More
Translated text
Key words
HTS-SQUID, ramp-edge Josephson junctions, robot-based NDE system, unshielded environment