A miniaturised sensor for continuous lactate measurement in saliva was developed and tested. The sensor was built using silicon microfabrication technologies. The size of the chip is 5.5mm×6.4mm×0.7mm and features a working, a counter and an Iridium reference electrode. The chip has a cavity whose floor is perforated by fine pores. The cavity contains the enzyme lactate oxidase (LOD), which is immobilised in an agarose gel. Prior to the amperometric detection of the reaction product hydrogen peroxide at the working electrode, the analyte lactate has to pass the pores to reach the cavity with the lactate oxidase by diffusion.
更多
查看译文
关键词
Lactate sensor,Saliva,Amperometric measurement,Silicon enzyme sensor,Iridium reference electrode