Temperature sensitivity of surface channel effects on high-purity germanium detectors

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(1995)

引用 27|浏览11
暂无评分
摘要
The temperature sensitivity of surface channel effects on planar high-purity germanium detectors was measured using 60-keV gamma-ray scanning techniques, as part of a radiation damage study. When measured in this manner, the surface effects on most detectors showed extreme temperature sensitivity in the 72–95 K region. The effect of the surface channel increased with increasing temperature to such an extent that the efficiency, as measured by the count rate in the 1332-keV peak from a 60Co source, decreased by a factor of over two in some cases. Since the peak efficiency for the 1332-keV gamma ray decreased as the temperature increased throughout the operating range (72–120 K) the effect of the surface channel must continue to increase beyond the temperature (95 K) at which the 60-keV scan loses its sensitivity because of the strong attenuation of these much lower energy gamma rays. Radiation damage had no measurable effect on the surface characteristics. No correlation between the surface effects and the resolution changes of the 1332-keV peak was observed.
更多
查看译文
关键词
gamma ray,radiation damage