Bioluminescent flow sensor for L-phenylalanine determination in serum.

TALANTA(1993)

Cited 14|Views13
No score
Abstract
A highly sensitive and rapid bioluminescent flow sensor was developed for the determination of the content of L-phenylalanine (Phe) in serum by monitoring the reduced form of nicotinamide adenine dinucleotide (NADH), produced by immobilized phenylalanine dehydrogenase (PheDH), with bacterial bioluminescent enzymes immobilized on a separate nylon coil. The L-PheDHs extracted from Bacillus badius, Bacillus sphaericus and Rhodococcus sp. M4 were investigated and the performances of the three immobilized L-PheDH's were analysed. The B. badius reactor was found to give higher transformation rate and better sensitivity; the response was linear from 1 to 100muM at 25-degrees, with a detection limit of 10 pmoles (0.5muM). The intra- and inter-assay coefficients of variation were less than 5% and recoveries ranged from 90 to 101%. The results agreed well with those obtained with a chromatographic method for the Phe determination in serum and with the normal reference values.
More
Translated text
Key words
bioluminescent flow sensor,serum,l-phenylalanine