谷歌浏览器插件
订阅小程序
在清言上使用

Whole Blood Thallium Determination by GFAAS with High-Frequency Modulation Polarization Zeeman Effect Background Correction

Biological Trace Element Research(2010)

引用 14|浏览1
暂无评分
摘要
A new technique of blood thallium direct determination based on graphite furnace atomic absorption spectrometry with Zeeman background absorption correction system was designed. The developed technique does not require sample digestion. Sample treatment includes only a fivefold per volume dilution of blood sample with 0.1% ( m / v ) Triton X-100. L’vov integrated platform was modified with 400 μg of Rh. Matrix modifier (200 μg NH 4 NO 3 and 160 μg Pd(NO 3 ) 2 ) was suggested for coping chloride and blood organic matter interferences. Standard reference material (Clincheck® Plasma Control for trace elements) analysis was used for validation. Additional validation was performed by analyzing spiked blood samples in the whole dynamic range. The dynamic range was 2–50 μg/L. Precision (RSD) was found <12%. Blood thallium limit of detection was 0.2 μg/L.
更多
查看译文
关键词
Trace element,Blood analysis,Thallium,Graphite furnace atomic absorption spectrometry,Modifiers in GFAAS
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper