Estudo biomecânico da rigidez da osteossíntese com placas em ponte em tíbias de cadáveres humanos

Acta Ortopedica Brasileira(2010)

引用 2|浏览11
暂无评分
摘要
Objective To compare the stiffness of three different assemblies of bridging plates with intramedullary locking nails in cadaveric models of tibial fractures type C Materials and Methods Twenty cadaveric tibias subjected to type C fractures, fifteen were fixed with the bridging plate technique and divided into three groups, according to the plate size (10, 14 and 18 holes), and five were fixed with intramedullary nail All the tibias were exposed to similar and progressive loads Dislocation of both fragments (proximal and distal) was measured on three planes (sagittal, coronal and axial), as the load was increased Results tibias fixed with the 18 hole bridging plate have the same biomechanical behavior of tibias fixed with intramedullary locking nails Conclusions In type C tibial fractures, there is more mobility at the distal segment on the coronal plane when the fracture is fixed with 14 and 18 holes bridging plates as compared to fractures fixed with non-reamed intramedullary looking nails Even with this mobility, the relative movements between the fragments at GHB and GP18 seem to be similar
更多
查看译文
关键词
Tibial fractures,Bone plates,Internal fixation,Bone nails