Multi-wavelength calibration procedure for the pierre Auger Observatory Fluorescence Detectors

Astroparticle Physics(2009)

Cited 11|Views8
No score
Abstract
We present a method to measure the relative spectral response of the Pierre Auger Observatory Fluorescence Detector. The calibration was done at wavelengths of 320, 337, 355, 380 and 405nm using an end-to-end technique in which the response of all detector components are combined in a single measurement. A xenon flasher and notch-filters were used as the light source for the calibration device. The overall uncertainty is 5%.