Reproducibility Of Multi-Slice Spiral Computed Tomography Scans: An Experimental Study

MEDICAL PHYSICS(2004)

引用 7|浏览25
暂无评分
摘要
In multi-slice spiral computed tomography (CT) images interpolation artifacts are present. The relationship between the x-ray tube rotation angle and these artifacts is demonstrated. A head phantom was repeatedly scanned with a four-slice CT scanner at different pitch values. Two scans, made with identical scan parameters, nearly always have different x-ray tube starting angles. Consequently, artifacts appeared differently and residual artifacts appeared when subtractions were made. We conclude that only if the x-ray tube starting angle is equal for both scans or if a very low pitch is used, images are highly reproducible. (C) 2004 American Association of Physicists in Medicine.