Stentplazierung unter Echtzeit-MR-Kontrolle: erste tierexperimentelle Erfahrungen

A. Bücker,J. Neuerburg,G. Adam, K. Schürmann,V. Rasche,J. Vaals, A. Molgaard-Nielsen,R. Günther

ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN(1998)

引用 33|浏览9
暂无评分
摘要
Purpose: The aim of this study was to test the feasibility of iliac artery stent placement under MR guidance with real-time MR radial scanning in an animal model. Materials and Methods: The experiments were performed on three pigs in a 1.5 T scanner. Radial scanning with a gradient echo technique (T-R 8.4 ms, T-E 3.6 ms, flip angle 10 degrees) was used. A dedicated backprojector performed the reconstruction of the raw data in real-time. The resulting MR-images were displayed on LCD screens beside the magnet. The sliding window reconstruction technique allowed image acquisition at a frame rate of 16 images per second. MR-compatible self-expanding stents with a diameter of 8 mm and a length of 3 cm were placed into the left iliac artery. Their positions were verified by digital subtraction angiography (DSA) and compared to MRI. Results: All stents were successfully placed. Stent positions as monitored by real-time MR were identical to those seen on DSA images. The time needed for exact positioning of the scan plane ranged from 15 to 30 minutes. Stent placement itself took 8 minutes on average. Conclusion: Radial scanning applied together with the sliding window reconstruction technique allows placement of stents in iliac arteries under realtime MR control.
更多
查看译文
关键词
intervention, MR-guided,real-time MR,radial scanning,stents