Parametric evaluation of laser ablation and ionization time-of-flight mass spectrometry with ion guide cooling cell

Spectrochimica Acta Part B: Atomic Spectroscopy(2008)

Cited 24|Views5
No score
Abstract
A novel laser ablation and ionization time-of-flight mass spectrometer has been used for direct elemental analysis of alloys. The system was incorporated with an ion guide cooling cell to reduce the kinetic energy distribution for the purpose of better resolution. Parametric studies have been conducted on the system with respect to the buffer gas pressure and the distance from sample to the nozzle to obtain the maximal signal intensities. In order to obtain satisfactory relative sensitivity coefficients (RSC) for different elements, the influence of the laser irradiance, nozzle voltage, rf frequency and voltage of the hexapole were also investigated. Under the optimized conditions, the RSC of different elements were available for direct semi-quantitative analysis. The mass resolving power (FWHM) of the spectrometer was approximately 7000 (m/Δm) and the limit of detection (LOD) was 10−6 g/g.
More
Translated text
Key words
Laser ionization,Mass spectrometry,Elemental analysis,Direct solids analysis
AI Read Science