Cycle testing of the MagScrew total artificial heart external battery pack: update I.

ARTIFICIAL ORGANS(2011)

Cited 2|Views5
No score
Abstract
MagScrew total artificial heart (TAH) external battery pack (EBP) cycle bench testing continued over a period of 18 months using two fresh Wilson Greatbatch lithium ion EBPs during continuous charge and discharge cycles under a simulated TAH system current requirement. The same electronic load developed for our initial testing was used to simulate the MagScrew current waveforms typically observed during nominal operation. The current load profiles for this test were modified from the ones previously described and applied to the EBP under test during a voltage-defined discharge cycle. The test ended when EBP#2 reached end of life at 1450 cycles. At that point, EBP#1 remained healthy with a capacity of 175 min until full discharge. Performance of EBP#2 was still within expected ranges. Performance of EBP#1 exceeded expectations. These differences are probably caused by slight manufacturing changes. More tests will provide additional data to define a statistical distribution to better characterize EBP performance. In conclusion, endurance performance of the EBP remained satisfactory.