Photomultiplier nonlinear response in time-domain laser-induced luminescence spectroscopy

Leandro José Bossy Schip, Bruno Phelippe Buzelatto,Fábio Roberto Batista,Carlos Jorge da Cunha, Lauro Camargo Dias Jr.,João Batista Marques Novo

QUIMICA NOVA(2007)

Cited 3|Views7
No score
Abstract
A new procedure to find the limiting range of the photomultiplier linear response of a low-cost, digital oscilloscope-based time-resolved laser-induced luminescence spectrometer (TRLS), is presented. A systematic investigation on the instrument response function with different signal input terminations, and the relationship between the luminescence intensity reaching the photomultiplier and the measured decay time are described. These investigations establish that setting the maximum intensity of the luminescence signal below 0.3V guarantees, for signal input terminations equal or higher than 99.7 ohm, a linear photomultiplier response.
More
Translated text
Key words
photomultiplier,time-resolved,laser-induced luminescence